Welcome to the Warner Instruments Knowledge Base
Please look around
We've structured this KB to mirror our product website. However, if you can't find your answer, then please submit a question and we'll respond directly to you.
|
Search the Knowledgebase |
Browse by Category |
|
|
|
View Questions by Category |
|
|
Chamber Accessories
|
There were 11 questions found in this category:
Can Warner Series 20 Perfusion Chambers be used with #0 thickness Cover Glass
Nothing speaks against the use of #0 thickness glass coverslips with our Series 20 chambers. However, you might want to use a P- or PH- platform with screw-clamps rather than a PM- platform with magnetic clamps. The screw- clamps will allow you to adjust the clamp pressure. Be careful not to br ...
How do I replace the thermistor on a TA-29 cable?
1. Remove insulating tubing covering solder joints between TS-70B thermistor and shielded cable.2. Remove inner insulating tubing covering solder joints between TS-70B thermistor and inner-conductor of shielded cable.3. Unsolder broken thermistor and solder on replacement thermistor.4. Replace i ...
The prongs on my platform heater broke. Are replacements available?
Yes, the RH-2 Replacement Heater Elements for Series 20 Chambers are the desired replacement parts. They come in packs of 2 and the order number is 64-0274.
What are the inner and outer diameters of the various sizes of polyethylene tubing Warner sells?
Tubing Model Part Number Inner Diameter Outer Diameter PE-10 640750 0.28mm 0.61mm PE-50 640752 0.58mm 0.97mm PE-90 640754 0.86mm 1.27mm PE-160 640755 1.14mm 1.57mm
What type of grease is used for sealing the gaskets and coverglass on RC Series chambers?
The grease we recommend is the Dow 111 silicone grease that is normally used with vacuum pump tubing line connections. It is also used to seal glass dessicators. We sell two products that contain a tube of this grease. Order number 64-0275 is simply a tube of Dow 111 silicone, while order number ...
What type of tubing does the RC-26GLP chamber use?
The flexible tubing supplied with the RC-26GLP is PE-160 (64-0755, 10 feet). The standard, supplied stainless steel suction tube is the ST-1R (64-1400). Alternatively you have the option of using the ST-3R suction tube (64-1407), which has the new micro-slit designBoth suction tubes are compatib ...
Which mesh sizes are offered with slice anchors?
All kits come with nylon mesh in two sizes (112 micron and 160 micron) that may be cut and used in conjunction with the slice anchor. Each sheet is approximately 7x7 cm.
Why is my TA-29 bead thermistor recording innaccurate temperatures?
It's possible that the lead wire is broken, in which case you can purchase a TS-70B replacement part (part # 640270) and solder it to the existing cable.Often people mistakenly plug the TA-29 into the T1 or T2 output BNC connectors on the front of the temperature controller. The TA-29 needs to b ...
Are coverslips appropriate for cell culturing?
Coverslips are appropriate for cell culture but are not coated. They will, however, accept Cell-Tak and poly-L coatings if needed.
Don't see your answer? Please submit a question!
We welcome your questions! Submit a question to us (via email) by clicking here and we'll answer you directly. We'll also anonymously post your question to our FAQ Knowledge Base (KB) to assist other people.If you choose to not communicate via this KB, then please send your questions directly to ...
What is the thickness of the coverglass you carry?
We carry coverglass in three thicknesses: #1, #1.5, and #2Generally speaking, glass thickness is expressed as a range to allow for manufacturing variance and our glass adheres to industry standards.#1 = 0.13 - 0.17 mm (130-170 µm)#1.5 = 0.16 - 0.19 mm (160-190 µm) #2 = 0.17 - 0.25 mm ...
|
|