There were 7 questions found in this category:
  1. questionCan the IE-210 headstages be repaired?
    No, it is not cost effective to repair an IE-210 headstage. It is less expensive to replace it than to repair it.

  2. questionCan the intracellular electrometers (IE-210, IE-251A) be used to study the blood vessel of rat?
    The IE-210 can be used with any cell type that carries a membrane potential or charge. The amplifier is essentially a current clamp with added bridge balance circuitry for current injection.

  3. questionCan the leads in the headstage DIN connector be repaired?
    Yes. While we can make a repair at the factory, it is a simple enough task that the user can make a repair on-site.The connector used for most Warner devices (amplifiers, temperature controllers, etc.) are known as DIN type connectors. They can be opened by removing the set-screw on the side and ...

  4. questionIs there a way to avoid damaging the IE-210 intracellular amplifier probe/headstage?
    The probe input is not protected against electrostatic discharge, even if the amplifier is in the OFF mode. Therefore, make sure to electrically discharge yourself through a connection to ground before touching the probe.

  5. questionWhat are the differences between chassis and circuit ground?
    Chassis ground is used to connect the instrument case, or chassis, to an earth ground. It is not usually necessary to use this ground point, and not using it can avoid the occurrence of 'ground loops'.Circuit ground is a reference point chosen for the instrument circuit voltages. This reference ...

  6. questionWhat is the VA (wattage) of the IE-210?
    The VA is 20 watts.

  7. questionDon't see your answer? Please submit a question!
    We welcome your questions! Submit a question to us (via email) by clicking here and we'll answer you directly. We'll also anonymously post your question to our FAQ Knowledge Base (KB) to assist other people.If you choose to not communicate via this KB, then please send your questions directly to ...